General description
Spectrometer is dedicated to FTIR analysis of biological (aqueous) samples using two dedicated measurement cells: AquaSpec – transmittance cell with optical pathlength of ~ 7 µm and BioATRII, Attenuated Total Reflection cell. Measurements are performed in precisely regulated temperature 4 – 98 °C. Dedicated OPUS 7.0 software is used for data collection and analysis. Automatic measurements at temperatures ramps are possible.
Applications
- Protein secondary structure analysis. By analysis of amide I band (1600 -1800 cm-1) shape in FTIR spectrum, that corresponds to C=O stretching vibration within peptide bonds it is possible to measure secondary structure content of protein in a sample.
- Protein stability analysis. Spectra are recorded in increasing temperature and therefore changes in protein secondary structure induced by heating can be assessed.